Absolute Dosimetry


Absolute Dosimetry

The following table is a guide to select the appropriate detector:

Farmer type cylindrical chambers for Absolute Dosimetry of photon and electron beams in air, solid or water.
  FC65-G FC65-P FC23-C  
Active volume (cm3)
0.65 0.65 0.23  
Outer electrode (g/cm3)
Graphite
1.82
POM
1.42
C552
1.76
 
Inner electrode (g/cm3)
Aluminium
2.7
Aluminium
2.7
C552
1.76
 
Sensitivity (Gy/C x 107)
5 5 14  


Parallel plate chambers for Absolute Dosimetry of electron beams
  PPC05 PPC40 NACP  
Inner electrode diameter (mm)
10 16 10  
Outer diameter (mm)
30 44 30  
Guarding width (mm)
1 4 3  
Sensitivity (Gy/C x 107)
58 9 16  


Application energy range for photons from Co60 - 50 MV and for electrons from 4 MeV - 50 MeV. 
 
In principal, cylindrical chambers with a volume >/= 0.1 cm3 can be used for Absolute Dosimetry.

 

 